内容記述 | The need for passive microwave radiometry using aperture synthesis and phased arrays for large apertures requiring more stability has generated several studies. The in-flight calibration and, therefore, the stability of such systems is an important design consideration. A 4.3 GHz laboratory radiometry operating in a balanced Dicke mode has been developed for making precise brightness (noise) temperature measurements. The stability of the noise diode source used in the noise-injection loop is of interest. A statistical Allan variance method to characterize the stability of the radiometer and the noise diode is presented in this paper. Noise measurements which demonstrate this approach are also included. |