| タイトル | A new method for monitoring long term calibration of the SBUV and TOMS instruments |
| 本文(外部サイト) | http://hdl.handle.net/2060/19950004705 |
| 著者(英) | Ahmad, Z.; Wellemeyer, C.; Seftor, C. |
| 著者所属(英) | Hughes STX, Inc. |
| 発行日 | 1994-04-01 |
| 言語 | eng |
| 内容記述 | A new method has been developed to monitor the long-term calibration of the Solar Backscatter Ultraviolet (SBUV) and Total Ozone Mapping Spectrometer (TOMS) instruments. It is based on the fact that the radiance in one channel can be expressed as a linear sum of the radiances in neighboring channels. Using simulated radiances for the SBUV and TOMS instruments, various scenarios of changes in instrument calibration are investigated. Results from sample processing of SBUV data are also presented. |
| NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
| レポートNO | 95N11118 |
| 権利 | No Copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/110582 |