| タイトル | Nonintrusive fast response oxygen monitoring system for high temperature flows |
| 本文(外部サイト) | http://hdl.handle.net/2060/19940013182 |
| 著者(英) | Stanton, Alan C.; Oh, Daniel B. |
| 著者所属(英) | Southwest Sciences, Inc. |
| 発行日 | 1993-10-01 |
| 言語 | eng |
| 内容記述 | A new technique has been developed for nonintrusive in situ measurement of oxygen concentration, gas temperature, and flow velocity of the test media in hypersonic wind tunnels. It is based on absorption of near-infrared radiation from inexpensive GaAlAs laser diodes used in optoelectronics industry. It is designed for simultaneous measurements along multiple lines of sight accessed by fiber optics. Molecular oxygen concentration is measured from the magnitude of absorption signals; rotational gas temperature is measured from the intensity ratio of two oxygen absorption lines; and the flow velocity is measured from the Doppler shift of the absorption line positions. This report describes the results of an extensive series of tests of the prototype instrument in laboratory flames emphasizing assessment of the instruments capabilities for quantitative measurement of O2 concentration (mole fraction) and gas temperature. |
| NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
| レポートNO | 94N17655 NASA-CR-4553 NAS 1.26:4553 |
| 権利 | No Copyright |