JAXA Repository / AIREX 未来へ続く、宙(そら)への英知

このアイテムに関連するファイルはありません。

タイトルECC ozonesonde performance at high altitudes: Pump efficiency
本文(外部サイト)http://hdl.handle.net/2060/19810017068
著者(英)Torres, A. L.
著者所属(英)NASA Wallops Flight Center
発行日1981-05-01
言語eng
内容記述The ECC ozonesonde sampling behavior was examined at pressures ranging from 60 to 6 hPa with the objective of evaluating uncertainties in high altitude ozone data caused by variations in pumping efficiency. The averaged pump efficiency correction curve for a 43 sample set of 3A type ECC pumps showed a 2-3% bias from the curve provided by the manufacturer. In addition, random pump to pump variations (2 sigma) + or - were 5% at 6 hPa. These values probably represent minimum errors since the pumps were from the same production batch. A seven sample set of the newer 4A type ECC pumps was examined, with similar findings as for the 3A types.
NASA分類GEOPHYSICS
レポートNO81N25604
NASA-TM-73290
権利No Copyright
URIhttps://repository.exst.jaxa.jp/dspace/handle/a-is/168814


このリポジトリに保管されているアイテムは、他に指定されている場合を除き、著作権により保護されています。