| タイトル | ECC ozonesonde performance at high altitudes: Pump efficiency |
| 本文(外部サイト) | http://hdl.handle.net/2060/19810017068 |
| 著者(英) | Torres, A. L. |
| 著者所属(英) | NASA Wallops Flight Center |
| 発行日 | 1981-05-01 |
| 言語 | eng |
| 内容記述 | The ECC ozonesonde sampling behavior was examined at pressures ranging from 60 to 6 hPa with the objective of evaluating uncertainties in high altitude ozone data caused by variations in pumping efficiency. The averaged pump efficiency correction curve for a 43 sample set of 3A type ECC pumps showed a 2-3% bias from the curve provided by the manufacturer. In addition, random pump to pump variations (2 sigma) + or - were 5% at 6 hPa. These values probably represent minimum errors since the pumps were from the same production batch. A seven sample set of the newer 4A type ECC pumps was examined, with similar findings as for the 3A types. |
| NASA分類 | GEOPHYSICS |
| レポートNO | 81N25604 NASA-TM-73290 |
| 権利 | No Copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/168814 |