タイトル | Recovery and radiation corrections and time constants of several sizes of shielded and unshielded thermocouple probes for measuring gas temperature |
本文(外部サイト) | http://hdl.handle.net/2060/19780007520 |
著者(英) | Glawe, G. E.; Holanda, R.; Krause, L. N. |
著者所属(英) | NASA Lewis Research Center |
発行日 | 1978-01-01 |
言語 | eng |
内容記述 | Performance characteristics were experimentally determined for several sizes of a shielded and unshielded thermocouple probe design. The probes are of swaged construction and were made of type K wire with a stainless steel sheath and shield and MgO insulation. The wire sizes ranged from 0.03- to 1.02-mm diameter for the unshielded design and from 0.16- to 0.81-mm diameter for the shielded design. The probes were tested through a Mach number range of 0.2 to 0.9, through a temperature range of room ambient to 1420 K, and through a total-pressure range of 0.03 to 0.2.2 MPa (0.3 to 22 atm). Tables and graphs are presented to aid in selecting a particular type and size. Recovery corrections, radiation corrections, and time constants were determined. |
NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
レポートNO | 78N15463 E-9289 NASA-TP-1099 |