タイトル | A Portable Infrasonic Detection System |
本文(外部サイト) | http://hdl.handle.net/2060/20080034649 |
著者(英) | Burkett, Cecil G.; Lawrenson, Christopher C.; Zuckerwar, Allan J.; Masterman, Michael; Shams, Qamar A. |
著者所属(英) | NASA Langley Research Center |
発行日 | 2008-08-19 |
言語 | eng |
内容記述 | During last couple of years, NASA Langley has designed and developed a portable infrasonic detection system which can be used to make useful infrasound measurements at a location where it was not possible previously. The system comprises an electret condenser microphone, having a 3-inch membrane diameter, and a small, compact windscreen. Electret-based technology offers the lowest possible background noise, because Johnson noise generated in the supporting electronics (preamplifier) is minimized. The microphone features a high membrane compliance with a large backchamber volume, a prepolarized backplane and a high impedance preamplifier located inside the backchamber. The windscreen, based on the high transmission coefficient of infrasound through matter, is made of a material having a low acoustic impedance and sufficiently thick wall to insure structural stability. Close-cell polyurethane foam has been found to serve the purpose well. In the proposed test, test parameters will be sensitivity, background noise, signal fidelity (harmonic distortion), and temporal stability. The design and results of the compact system, based upon laboratory and field experiments, will be presented. |
NASA分類 | Acoustics |
権利 | Copyright, Distribution as joint owner in the copyright |
|