| タイトル | Wearable Wireless Telemetry System for Implantable Bio-MEMS Sensors |
| 本文(外部サイト) | http://hdl.handle.net/2060/20060051726 |
| 著者(英) | Simons, Renita E.; Miranda, Felix A.; Wilson, Jeffrey D.; Simons, Rainee N. |
| 著者所属(英) | NASA Glenn Research Center |
| 発行日 | 2006-10-01 |
| 言語 | eng |
| 内容記述 | In this paper, a telemetry and contact-less powering system consisting of an implantable bio-MEMS sensor with a miniature printed square spiral chip antenna and an external wearable garment with printed loop antenna is investigated. The wearable garment pick-up antenna and the implantable chip antenna are in close proximity to each other and hence couple inductively through their near-fields and behave as the primary and the secondary circuits of a transformer, respectively. The numerical and experimental results are graphically presented, and include the design parameter values as a function of the geometry, the relative RF magnetic near-field intensity as a function of the distance and angle, and the current density on the strip conductors, for the implantable chip antenna. |
| NASA分類 | Electronics and Electrical Engineering |
| レポートNO | NASA/TM-2006-214422 E-15558 |
| 権利 | Copyright, Distribution as joint owner in the copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/259427 |