| タイトル | Hygrometer for Detecting Water in Partially Enclosed Volumes |
| 本文(外部サイト) | http://hdl.handle.net/2060/20110016449 |
| 著者(英) | Youngquist, Robert C.; Parks, Steve; Surma, Jan |
| 著者所属(英) | NASA Kennedy Space Center |
| 発行日 | 2005-12-01 |
| 言語 | eng |
| 内容記述 | A portable hygrometer has been devised to implement a pre-existing technique for detecting water trapped in partially enclosed volumes that may be difficult to reach and cannot be examined directly. The technique is based on the fact that eventually the air in such a volume becomes saturated or nearly so. The technique is straightforward: One measures the relative humidity and temperature of both the ambient air and a sample of air from the enclosed volume. If the relative humidity of the sample is significantly greater than that of the ambient air and/or if the sample is at or close to the dew point, then it can be concluded that water is trapped in the volume. Of course, the success of this technique depends on the existence of an access hole through which one can withdraw some air from the enclosed volume. |
| NASA分類 | Man/System Technology and Life Support |
| レポートNO | KSC-12593 |
| 権利 | Copyright, Distribution as joint owner in the copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/269265 |