タイトルDetermining blood and plasma volumes using bioelectrical response spectroscopy
著者(英)Suire, S. S.; Leig, J.; Siconolfi, S. F.; Nusynowitz, M. L.; Moore, A. D. Jr
発行日1996-12-01
言語eng
内容記述We hypothesized that an electric field (inductance) produced by charged blood components passing through the many branches of arteries and veins could assess total blood volume (TBV) or plasma volume (PV). Individual (N = 29) electrical circuits (inductors, two resistors, and a capacitor) were determined from bioelectrical response spectroscopy (BERS) using a Hewlett Packard 4284A Precision LCR Meter. Inductance, capacitance, and resistance from the circuits of 19 subjects modeled TBV (sum of PV and computed red cell volume) and PV (based on 125I-albumin). Each model (N = 10, cross validation group) had good validity based on 1) mean differences (-2.3 to 1.5%) between the methods that were not significant and less than the propagated errors (+/- 5.2% for TBV and PV), 2) high correlations (r > 0.92) with low SEE (< 7.7%) between dilution and BERS assessments, and 3) Bland-Altman pairwise comparisons that indicated "clinical equivalency" between the methods. Given the limitation of this study (10 validity subjects), we concluded that BERS models accurately assessed TBV and PV. Further evaluations of the models' validities are needed before they are used in clinical or research settings.
NASA分類Life Sciences (General)