タイトル | Response of a tissue equivalent proportional counter to neutrons |
著者(英) | Gibbons, F.; Robbins, D. E.; Badhwar, G. D.; Braby, L. A. |
発行日 | 2002-12-01 |
言語 | eng |
内容記述 | The absorbed dose as a function of lineal energy was measured at the CERN-EC Reference-field Facility (CERF) using a 512-channel tissue equivalent proportional counter (TEPC), and neutron dose equivalent response evaluated. Although there are some differences, the measured dose equivalent is in agreement with that measured by the 16-channel HANDI tissue equivalent counter. Comparison of TEPC measurements with those made by a silicon solid-state detector for low linear energy transfer particles produced by the same beam, is presented. The measurements show that about 4% of dose equivalent is delivered by particles heavier than protons generated in the conducting tissue equivalent plastic. c2002 Elsevier Science Ltd. All rights reserved. |
NASA分類 | Life Sciences (General) |