タイトル | Photometer calibration error using extended standard sources |
著者(英) | Kennedy, B. C.; Torr, D. G.; Hays, P. B.; Torr, M. R. |
著者所属(英) | NASA Goddard Space Flight Center |
発行日 | 1976-03-01 |
言語 | eng |
内容記述 | As part of a project to compare measurements of the night airglow made by the visible airglow experiment on the Atmospheric Explorer-C satellite, the standard light sources of several airglow observatories were compared with the standard source used in the absolute calibration of the satellite photometer. In the course of the comparison, it has been found that serious calibration errors (up to a factor of two) can arise when a calibration source with a reflecting surface is placed close to an interference filter. For reliable absolute calibration, the source should be located at a distance of at least five filter radii from the interference filter. |
NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
レポートNO | 92A38106 |
権利 | Copyright |
URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/327212 |