タイトルHydrazine Detection with a Tunable Diode Laser Spectrometer
著者(英)Houseman, John; Brown, Pamela R.; Margolis, J. S.; Jackson, Julie R.; May, R. D.; Anderson, M. S.; Webster, C. R.
著者所属(英)Jet Propulsion Lab., California Inst. of Tech.
発行日1999-04-01
言語eng
内容記述Several instruments have been developed to measure low concentrations of hydrazine but none completely meet the sensitivity requirements while satisfying additional criteria such as quick response, stable calibration, interference free operation, online operation, reasonable cost, etc. A brief review is presented of the current technology including the electrochemical cell, the ion mobility spectrometer, the mass spectrometer, and the gas chromatograph. A review of the advantages and disadvantages of these instruments are presented here. The review also includes commercially unavailable technology such as the electronic nose and the Tunable Diode Laser (TDL) IR Spectrometer. It was found that the TDL could meet the majority of these criteria including fast response, minimum maintenance, portability, and reasonable cost. An experiment was conducted to demonstrate the feasibility of such a system using an existing (non-portable) instrument. A lead-salt tunable diode laser, cooled to 85 degrees Kelvin was used to record direct absorption and second-derivative spectra of Hydrazine at several pressures to study the sensitivity to low levels of Hydrazine. Spectra of NH3 and CO2 were used for wavelength identification of the scanned region. With a pathlength of 80 m, detection sensitivities of about 1 ppb were achieved for hydrazine in dry nitrogen at a cell pressure of 100 mbar. For spectroscopic detection of Hydrazine, spectral regions including strong Ammonia or Carbon Dioxide lines must be avoided. Strong Hydrazine absorption features were identified at 940/cm showing minimal contribution from Ammonia interferences as suitable candidates for Hydrazine gas detection. For the studies reported here, the particular laser diode could only cover the narrow regions near 962/cm and 965/cm where strong Ammonia interferences were expected. However, the high resolution (0.001/cm) of the TDL spectrometer allowed individual lines of Hydrazine to be identified away from interferences from either Ammonia or Carbon Dioxide, especially at lower pressures. A Hydrazine line was identified at 961.75/cm which was free from Ammonia absorption and would be suitable to monitor hydrazine levels. This paper also shows data on the degradation of the detectivity of Hydrazine with increasing pressure. Several design options for a portable unit are presented, including designs with near IR and Quantum Cascade laser components which do not require liquid nitrogen cooling.
NASA分類Instrumentation and Photography
権利No Copyright


このリポジトリに保管されているアイテムは、他に指定されている場合を除き、著作権により保護されています。