| タイトル | New high-resolution electrostatic ion mass analyzer using time of flight |
| 著者(英) | Lundgren, R. A.; Sheldon, R. B.; Gloeckler, G.; Hamilton, D. C.; Ipavich, F. M. |
| 著者所属(英) | Maryland Univ. |
| 発行日 | 1990-10-01 |
| 言語 | eng |
| 内容記述 | The design of a high-resolution ion-mass analyzer is described, which is based on an accurate measurement of the time of flight (TOF) of ions within a region configured to produce a harmonic potential. In this device, the TOF, which is independent of ion energy, is determined from a start pulse from secondary electrons produced when the ion passes through a thin carbon foil at the entrance of the TOF region and at a stop pulse from the ion striking a microchannel plate upon exciting the region. A laboratory prototype instrument called 'VMASS' was built and was tested at the Goddard Space Flight Center electrostatic accelerator, showing a good mass resolution of the instrument. Sensors of the VMASS type will form part of the WIND Solar Wind and Suprathermal Ion experiment, the Soho mission, and the Advanced Composition Explorer. |
| NASA分類 | SPACECRAFT INSTRUMENTATION |
| レポートNO | 91A12937 |
| 権利 | Copyright |