JAXA Repository / AIREX 未来へ続く、宙(そら)への英知

このアイテムに関連するファイルはありません。

タイトルResponse characteristics of laser diffraction particle size analyzers - Optical sample volume extent and lens effects
著者(英)Oechsle, V.; Hirleman, E. D.; Chigier, N. A.
著者所属(英)Carnegie-Mellon Univ.|Arizona State Univ.
発行日1984-10-01
言語eng
内容記述The response characteristics of laser diffraction particle sizing instruments were studied theoretically and experimentally. In particular, the extent of optical sample volume and the effects of receiving lens properties were investigated in detail. The experimental work was performed with a particle size analyzer using a calibration reticle containing a two-dimensional array of opaque circular disks on a glass substrate. The calibration slide simulated the forward-scattering characteristics of a Rosin-Rammler droplet size distribution. The reticle was analyzed with collection lenses of 63 mm, 100 mm, and 300 mm focal lengths using scattering inversion software that determined best-fit Rosin-Rammler size distribution parameters. The data differed from the predicted response for the reticle by about 10 percent. A set of calibration factor for the detector elements was determined that corrected for the nonideal response of the instrument. The response of the instrument was also measured as a function of reticle position, and the results confirmed a theoretical optical sample volume model presented here.
NASA分類INSTRUMENTATION AND PHOTOGRAPHY
レポートNO85A10037
権利Copyright
URIhttps://repository.exst.jaxa.jp/dspace/handle/a-is/394786