タイトル | Room-temperature mercuric iodide spectrometry for low-energy X-rays |
著者(英) | Economou, T. E.; Turkevich, A. L.; Dabrowski, A. J.; Barton, J. B.; Huth, G. C.; Iwanczyk, J. S.; Kusmiss, J. H. |
著者所属(英) | Chicago Univ.|University of Southern California |
発行日 | 1982-01-01 |
言語 | eng |
内容記述 | A discussion of the limits of energy resolution in different energy ranges is given. The energy resolution of a spectrometer is analyzed in terms of the parameters characterizing the crystal, the detector, and the amplification electronics. A high-resolution room-temperature HgI2 spectrometry system was used to measure low-energy X-ray fluorescence spectra. For the MgK-alpha X-ray line the measured resolution was 245 eV (fwhm); the electronic noise linewidth of the system was 225 eV. Alpha-particles were used to excite X-ray fluorescence from low-Z elements separately or in combination. The shape of the photopeaks in the spectra is discussed. |
NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
レポートNO | 83A36717 |
権利 | Copyright |
URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/403734 |
|