タイトルAn instrument for rapidly measuring plasma distribution functions with high resolution
著者(英)Carlson, C. W.; Curtis, D. W.; Michael, W.; Paschmann, G.
著者所属(英)Max-Planck-Inst. fuer Extraterrestrische Physik|California Univ.
発行日1982-01-01
言語eng
内容記述An instrument which can rapidly measure plasma particle distribution functions has been developed based upon recent innovations in electrostatic analyzer design and position sensitive particle detection. The new analyzer uses a quadrispherical geometry, but has a completely uniform 360 deg fan-shaped field of view. The polar angular distribution of entering particles is spatially imaged onto a position sensitive detector at the annular exit aperture after a deflection through 90 deg. Several methods of position sensitive detection have been successfully used in conjunction with this analyzer. The simplest is individual channel multipliers spaced around the annular exit. Microchannel plate electron multipliers permit greater position resolution to be obtained, and a detector using microchannel plates followed by a resistive anode image converter obtains angular resolution of about one degree, i.e., 360 individual angle pixels. Instruments of this type were flown on a sounding rocket in early 1982 and will be included on the Giotto comet mission and the AMPTE ion release module (IRM).
NASA分類SPACECRAFT INSTRUMENTATION