| タイトル | Calibration of a mass spectrometer experiment for ozone |
| 著者(英) | Anderson, S.; Mauersberger, K. |
| 著者所属(英) | Minnesota Univ. |
| 発行日 | 1981-07-01 |
| 言語 | eng |
| 内容記述 | A system with which to calibrate a stratospheric mass spectrometer experiment for ozone has been developed and evaluated. The stability of ozone and the formation of impurities within the calibration volume were measured. The half-life of ozone is about 12 hours, its purity ranges from 95% to 99% and is known to within 1%. The total calibration accuracy of the mass spectrometer experiment for ozone is about + or - 2%. |
| NASA分類 | SPACECRAFT INSTRUMENTATION |
| レポートNO | 81A42147 |
| 権利 | Copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/415788 |