JAXA Repository / AIREX 未来へ続く、宙(そら)への英知

このアイテムに関連するファイルはありません。

タイトルModified chemiluminescent NO analyzer accurately measures NOX
著者(英)Summers, R. L.
著者所属(英)NASA Lewis Research Center
発行日1978-06-01
言語eng
内容記述Installation of molybdenum nitric oxide (NO)-to-higher oxides of nitrogen (NOx) converter in chemiluminescent gas analyzer and use of air purge allow accurate measurements of NOx in exhaust gases containing as much as thirty percent carbon monoxide (CO). Measurements using conventional analyzer are highly inaccurate for NOx if as little as five percent CO is present. In modified analyzer, molybdenum has high tolerance to CO, and air purge substantially quenches NOx destruction. In test, modified chemiluminescent analyzer accurately measured NO and NOx concentrations for over 4 months with no denegration in performance.
NASA分類MATERIALS
レポートNO78B10047
LEW-12850
権利No Copyright
URIhttps://repository.exst.jaxa.jp/dspace/handle/a-is/436124