| タイトル | Variation of dose distribution by detectors for narrow beam |
| その他のタイトル | 細いビームに対する検出器による線量分布変化 |
| 著者(日) | 藤崎 達也; 齋藤 秀敏; 稲田 哲雄; 阿部 慎司; 福士 政広; 福田 賢一 |
| 著者(英) | Fujisaki, Tatsuya; Saito, Hidetoshi; Inada, Tetsuo; Abe, Shinji; Fukushi, Masahiro; Fukuda, Kenichi |
| 著者所属(日) | 茨城県立医療大学; 東京都立保健科学大学; 茨城県立医療大学; 茨城県立医療大学; 東京都立保健科学大学; 東京都立保健科学大学 |
| 著者所属(英) | Ibaraki Prefectural University of Health Sciences; Tokyo Metropolitan University of Health Sciences; Ibaraki Prefectural University of Health Sciences; Ibaraki Prefectural University of Health Sciences; Tokyo Metropolitan University of Health Sciences; Tokyo Metropolitan University of Health Sciences |
| 発行日 | 2000-08-08 |
| 刊行物名 | Proceedings of the Second International Workshop on EGS Proceedings of the Second International Workshop on EGS |
| 開始ページ | 264 |
| 終了ページ | 271 |
| 刊行年月日 | 2000-08-08 |
| 言語 | eng |
| 抄録 | Percentage depth dose (PDD) is the essential parameter for the absorbed dose determination. Since precise measurement is required for the narrow beam especially, several kinds of detectors are commonly used to measure dose distribution for narrow beam. However, the variation of response caused by kinds of the detectors has not been investigated precisely. In this study, PDDs were measured using several kinds of detectors, such as ionization chambers, silicon diode detectors and films. Furthermore, the absorbed dose of these detection materials in water was calculated using Monte Carlo simulation. As a result, it was shown that the response of ionization chambers (air-gas) was higher than diode detectors (silicon) and a film (photoemulsion). Therefore, the depth dose curves changed by kinds of the detector (materials). |
| キーワード | radiation dose distribution; radiation detector; narrow beam; ionization chamber; silicon diode detector; EGS4; Monte Carlo simulation; water; research and development; 放射線量分布; 放射線検出器; 細いビーム; イオンチェンバー; シリコンダイオード検出器; EGS4; モンテカルロシミュレーション; 水; 研究開発 |
| 資料種別 | Conference Paper |
| SHI-NO | AA0029759032 |
| レポートNO | KEK-Proceedings-2000-20 |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/43858 |