| タイトル | Determination of the line shapes of atomic nitrogen resonance lines by magnetic scans |
| 著者(英) | Stone, E. J.; Kley, D.; Lawrence, G. M. |
| 著者所属(英) | National Oceanic and Atmospheric Administration|Colorado Univ. |
| 発行日 | 1976-11-01 |
| 言語 | eng |
| 内容記述 | A technique is given for calibrating an atomic nitrogen resonance lamp for use in determining column densities of atoms in specific states. A discharge lamp emitting the NI multiplets at 1200 A and 1493 A is studied by obtaining absorption by atoms in a magnetic field (0-2.5 T). This magnetic scanning technique enables the determination of the absorbing atom column density, and an empirical curve of growth is obtained because the atomic f-value is known. Thus, the calibrated lamp can be used in the determination of atomic column densities. |
| NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
| レポートNO | 77A10809 |
| 権利 | Copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/440423 |