| タイトル | Low frost-point humidity generator |
| 著者(英) | Greenspan, L. |
| 発行日 | 1973-10-01 |
| 言語 | eng |
| 内容記述 | A low frost-point humidity generator has been developed at NBS to provide a capability for calibration, testing, and research at very low levels of water vapor content in such gases as atmospheric air, carbon dioxide and nitrogen. The generator produces frost points from -30 to -100 C at ambient pressures from 500 to 200,000 pascals (0.005 to 2 atm.). This is equivalent to mixing ratios of 4 micrograms to 51 grams of water vapor per kilogram of dry air and to vapor pressures of .0014 to 38 pascals. The generated test gas can be fed to a test chamber with independent temperature control between +25 and -100 C. The uncertainty of the frost point in the test chamber is estimated not to exceed 0.05 deg C. |
| NASA分類 | INSTRUMENTATION AND PHOTOGRAPHY |
| レポートNO | 74A24656 |
| 権利 | Copyright |
| URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/453401 |
|