タイトル | Heterodyne method for high specificity gas detection. |
著者(英) | Jaynes, D. N.; Williams, K. G.; Margozzi, A. P.; Donaldson, R. W.; Mcclatchie, E. A.; Gunter, W. D., Jr.; Dimeff, J.; Deboo, G. J. |
発行日 | 1971-11-01 |
言語 | eng |
内容記述 | This paper describes a new technique for measuring trace quantities of gases. The technique involves the use of a reference cell (containing a known amount of the gas being sought) and a sample cell (containing an unknown amount of the same gas) wherein the gas densities are modulated. Light passing through the two cells in sequence is modulated in intensity at the vibrational-rotational lines characteristic of the absorption spectrum for the gas of interest. Since the absorption process is nonlinear, modulating the two absorption cells at two different frequencies gives rise to a heterodyning effect, which in turn introduces sum and difference frequencies in the detected signal. Measuring the ratio of the difference frequency signal for example, to the signal introduced by the reference cell provides a normalized measure of the amount of the gas in the sample cell. The readings produced are thereby independent of source intensity, window transparency, and detector sensitivity. Experimental evaluation of the technique suggests that it should be applicable to a wide range of gases, that it should be able to reject spurious signals due to unwanted gases, and that it should be sensitive to concentrations of the order of 10 to the minus 8th power when used with a sample cell of only 20 cm length. |
NASA分類 | CHEMISTRY |
レポートNO | 72A10532 AIAA PAPER 71-1064 |
権利 | Copyright |
URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/464346 |
|