JAXA Repository / AIREX 未来へ続く、宙(そら)への英知

このアイテムに関連するファイルはありません。

タイトルCryogenic System for Interferometric Measurement of Dimensional Changes at 40 K: Design and Performance
本文(外部サイト)http://hdl.handle.net/2060/20070036804
著者(英)Crane, Allen; Madison, Tim; Miller, Franklin; Miller, David; Canavan, Edgar R.; Blake, Peter; Zukowski, Tim
著者所属(英)NASA Goddard Space Flight Center
発行日2007-01-01
言語eng
内容記述This report describes the facility, experimental methods, characterizations, and uncertainty analysis of the Cryo Distortion Measurement Facility (CDMF) at the Goddard Space Flight Center (GSFC). This facility is designed to measure thermal distortions of structural elements as the temperature is lowered from 320K to below 40 K over multiple cycles, and is capable of unattended running and data logging. The first measurement is to be the change in length and any bending of composite tubes with Invar end-fittings. The CDMF includes a chamber that is efficiently cooled with two cryo-coolers (one single-stage and one two-stage) rather than with liquid cryogens. Five optical ports incorporate sapphire radiation shields - transparent to the interferometer - on each of two shrouds and a fused silica vacuum-port window. The change in length of composite tubes is monitored continuously with displacement-measuring interferometers; and the rotations, bending, and twisting are measured intermittently with theodolites and a surface-figure interferometer. Nickel-coated invar mirrors and attachment mechanisms were developed and qualified by test in the CDMF. The uncertainty in measurement of length change of 0.4 m tubes is currently estimated at 0.9 micrometers.
NASA分類Mechanical Engineering
権利Copyright, Distribution as joint owner in the copyright


このリポジトリに保管されているアイテムは、他に指定されている場合を除き、著作権により保護されています。