タイトルMiniature metastable ionization detectors for exobiology flight experiments
著者(英)Woeller, F. H.
著者所属(英)NASA Ames Research Center
発行日1986-05-01
言語eng
内容記述The Metastable Ionization Detector (MID) is three orders of magnitude more sensitive than the thermal conductivity detectors used on previous flight instruments. The miniature MID provides scientists with a much smaller and highly sensitive detector for flight gas chromatographs. A miniature MID featuring an unconventional triaxial electrode configuration was developed and used routinely in the laboratory. Although much smaller and lighter than the commercial MID, its performance characteristics parallel those of the traditional design. The detector is compatible with the modulated voltage circuitry, also developed here, and thus can perform over an expanded response range of more than 7 orders magnitude. A micro volume version of a miniature MID, with an internal volume of less than 8 microliter, was recently designed is now being tested. The micro volume MID uses carrier gas flow rates of approx. 2cc/min thus eliminating the need for makeup gas when capillary columns are used.
NASA分類SPACE BIOLOGY
レポートNO86N26859
権利No Copyright
URIhttps://repository.exst.jaxa.jp/dspace/handle/a-is/529129