JAXA Repository / AIREX 未来へ続く、宙(そら)への英知

このアイテムに関連するファイルはありません。

タイトルA high resolution semiconductor detector for applications in space
本文(外部サイト)http://hdl.handle.net/2060/19720010048
著者(英)Shulman, H.; Alexander, P.
著者所属(英)Teledyne Isotopes
発行日1972-01-01
言語eng
内容記述Nuclear radiation detectors with volumes of approximately 1 cu cm was fabricated from single crystals of germanium-silicon alloy containing as much as 20 weight percent germanium. The properties of these detectors were investigated and will be discussed. Tests reveal that the gamma ray photoelectric peak efficiency of an alloy detector with only 12 weight percent germanium is approximately 4 times that of a silicon detector of equal volume. The room temperature roomure appears to be a good possibility. Storage for long periods at room temperature does not seem to adversely affect these devices. The results of preliminary radiation damage experiments suggest that the alloy detectors possess a radiation damage resistance far greater than that of silicon.
NASA分類INSTRUMENTATION AND PHOTOGRAPHY
レポートNO72N17698
権利No Copyright


このリポジトリに保管されているアイテムは、他に指定されている場合を除き、著作権により保護されています。