タイトルLong term variation on characteristics of germanium detectors
その他のタイトルゲルマニウム検出器特性の長期変化
著者(日)寺井 邦雄; 小村 和久
著者(英)Terai, Kunio; Komura, Kazuhisa
著者所属(日)島根県衛生公害研究所; 金沢大学 理学部 低レベル放射能実験施設
著者所属(英)Shimane Prefectural Institute for Public Health and Environmental Science; Kanazawa University Low Level Radioactivity Laboratory, Faculty of Science
発行日1995-01-17
刊行物名KEK Proceedings
KEK Proceedings
開始ページ233
終了ページ242
刊行年月日1995-01-17
言語eng
抄録Long term variation of characteristics of three germanium semiconductor detectors (Ge detector) were investigated. These detectors were two coaxial type lithium drifted Ge detectors with relative efficiency, 10.2 and 13.6 percent, respectively, and a coaxial type high purity Ge detector with its value 40.5 percent. Eight standard gamma ray calibration sources were used, and relative efficiency, output pulse width, peak to Compton ratio and liquid nitrogen consumption were checked for more than ten years. As a result, main causes of deterioration of Ge detector were found.
3個のゲルマニウム半導体検出器の特性変化を長期にわたって調べた。それらは、相対効率がそれぞれ10.2%と13.6%の同軸型リチウムドリフトゲルマニウム検出器、およびその値が、40.5%の同軸型高純度ゲルマニウム検出器である。較正線源として8種の標準γ線源を用い、10年以上の期間、液体窒素消費量、相対効率、出力パルス幅、ピーク対コンプトン比を調査した。その結果、性能低下の主な原因が明らかになった。
キーワードgermanium semiconductor detector; relative efficiency; peak to Compton ratio; lithium drifted germanium detector; high purity germanium detector; liquid nitrogen consumption; long term variation; nondestructive measurement; in situ measurement; calibration source; radiation detector; ゲルマニウム半導体検出器; 相対効率; ピーク対コンプトン比; リチウムドリフトゲルマニウム検出器; 高感度ゲルマニウム検出器; 液体窒素消費量; 長期変化; 非破壊測定; 現場測定; 較正線源; 放射線検出器