タイトル | The calculation of 3D saptial dose distribution around a shielded vaginal cylinder with iridium-192 source calculated by using Monte Carlo code EGS4 |
その他のタイトル | モンテカルロ・コードEGS4を用いて計算したIridium-192線源による遮蔽した鞘状円筒周囲の3次元線量位置分布の計算 |
著者(日) | Chen, I. J.; Sheu, R. D.; Chang, B. J.; Liu, Y. M.; Chao, L. S.; Yen, S. H. |
著者(英) | Chen, I. J.; Sheu, R. D.; Chang, B. J.; Liu, Y. M.; Chao, L. S.; Yen, S. H. |
著者所属(日) | Institute of Nuclear Energy Research; National Tsing Hua University; Institute of Nuclear Energy Research; Taipei Veterans General Hospital Cancer Therapy Center; Taipei Veterans General Hospital Cancer Therapy Center; Taipei Veterans General Hospital Cancer Therapy Center |
著者所属(英) | Institute of Nuclear Energy Research; National Tsing Hua University; Institute of Nuclear Energy Research; Taipei Veterans General Hospital Cancer Therapy Center; Taipei Veterans General Hospital Cancer Therapy Center; Taipei Veterans General Hospital Cancer Therapy Center |
発行日 | 2000-08-08 |
刊行物名 | Proceedings of the Second International Workshop on EGS Proceedings of the Second International Workshop on EGS |
開始ページ | 107 |
終了ページ | 114 |
刊行年月日 | 2000-08-08 |
言語 | eng |
抄録 | According to the previous study on the dose distribution around a shielded vaginal cylinder with a high-intensity Ir-192 source by using EGS4 code, the more detailed 3D-simulation model was established in this work. The results obtained by the 3D-simulation model were compared with those measured by ion chamber in previous work and evaluated by the existing treatment planning system. These results show good agreement. Furthermore, the 3D spatial dose distribution around a unit intensity Ir-192 source was calculated at each probable position in the vaginal cylinder with 90, 180, 270 deg tungsten shield and no shield cases. Those results save as a database which could be accessed through the graphical user interface developed by IDL(sup TM) to display dose distributions in 1D, 2D, and 3D plot under any combination of source intensity at each position, and shielding situation. It demonstrates a prototype of the brachytherapy treatment planning system based on the Monte Carlo EGS4 code. |
キーワード | three dimensional dose distribution; shielded vaginal cylinder; Ir-192 source; radiation therapy; Monte Carlo simulation; electron photon cascade; graphical user interface; research and development; EGS4; 3次元線量分布; 遮蔽した鞘状円筒; Ir-192線源; 放射線治療; モンテカルロシミュレーション; 電子・光子カスケード; グラフィック・ユーザー・インタフェース; 研究開発; EGS4 |
資料種別 | Conference Paper |
SHI-NO | AA0029759012 |
レポートNO | KEK-Proceedings-2000-20 |
URI | https://repository.exst.jaxa.jp/dspace/handle/a-is/55651 |