タイトル | Using a Cold Radiometer to Measure Heat Loads and Survey Heat Leaks |
本文(外部サイト) | http://hdl.handle.net/2060/20160008919 |
著者(英) | Hait, T.; Tuttle, J.; Dipirro, M.; Shirron, P. |
著者所属(英) | NASA Goddard Space Flight Center |
発行日 | 2014-01-29 |
言語 | eng |
内容記述 | We have developed an inexpensive cold radiometer for use in thermal/vacuum chambers to measure heat loads, characterize emissivity and specularity of surfaces and to survey areas to evaluate stray heat loads. We report here the results of two such tests for the James Webb Space Telescope to measure heat loads and effective emissivities of 2 major pieces of optical ground support equipment that will be used in upcoming thermal vacuum testing of the Telescope. |
NASA分類 | Spacecraft Design, Testing and Performance; Instrumentation and Photography |
レポートNO | GSFC-E-DAA-TN33587 |
権利 | No Copyright |