| タイトル | The Phoenix TECP Relative Humidity Sensor: Revised Results |
| 本文(外部サイト) | http://hdl.handle.net/2060/20140011327 |
| 著者(英) | Zent, Aaron |
| 著者所属(英) | NASA Ames Research Center |
| 発行日 | 2014-07-14 |
| 言語 | eng |
| 内容記述 | The original calibration function of the RH sensor on the Phoenix mission's Thermal and Electrical Conductivity Sensor (TECP), has been revised to correct the erroneously-published original calibration equation, to demonstrate the value of this unique data set, and to improve characterization of H2O exchange between the martian regolith and atmosphere. TECP returned two data streams, the temperature of the electronics analog board (Tb) and the digital 12-bit output of the RH sensor (DN), both of which are required to uniquely specify the H2O abundance. Because the original flight instrument calibration was performed against a pair of hygrometers that measured frost point (Tf), the revised calibration equation is also cast in terms of frost point. The choice of functional form for the calibration function is minimally constrained. A series of profiles across the calibration data cloud at constant DN and Tb does not reveal any evidence of a complex functional form. Therefore, a series of polynomials in both DN and Tb was investigated, along with several non-linear functions of DN and Tb. |
| NASA分類 | Astronomy |
| レポートNO | ARC-E-DAA-TN15703 |
| 権利 | No Copyright |