タイトル | Dual Transition Edge Sensor Bolometer for Enhanced Dynamic Range |
本文(外部サイト) | http://hdl.handle.net/2060/20040074282 |
著者(英) | Irwin, K. D.; Benford, D. J.; Chervenak, J. A.; Moseley, S. H. |
著者所属(英) | NASA Goddard Space Flight Center |
発行日 | 2004-04-01 |
言語 | eng |
内容記述 | Broadband surveys at the millimeter and submillimeter wavelengths will require bolometers that can reach new limits of sensitivity and also operate under high background conditions. To address this need, we present results on a dual transition edge sensor (TES) device with two operating modes: one for low background, ultrasensitive detection and one for high background, enhanced dynamic range detection. The device consists of a detector element with two transition temperatures (T(sub c)) of 0.25 and 0.51 K located on the same micromachined, thermally isolated membrane structure. It can be biased on either transition, and features phonon-limited noise performance at the lower T(sub c). We measure noise performance on the lower transition 7 x 10(exp -18) W/rt(Hz) and the bias power on the upper transition of 12.5 pW, giving a factor of 10 enhancement of the dynamic range for the device. We discuss the biasable range of this type of device and present a design concept to optimize utility of the device. |
NASA分類 | Instrumentation and Photography |
権利 | No Copyright |
|